Advances in Methodologies for Point-of-Care Testing of Coagulation Parameters

Xinyi Wang*
Guangdong Country Garden School

* Corresponding author: [email protected]

Abstract

Prothrombin Time (PT) and Activated Partial Thromboplastin Time (APTT) are utilized to assess the time required for the extrinsic and intrinsic pathways of blood coagulation, respectively, and are commonly used to monitor dosages of anticoagulant medications. Point-of-care testing (POCT) methodologies for coagulation primarily include electrochemical, flow-based, viscoelastic, and optical methods. This paper compares these methods from the perspectives of principle, application, and user experience, elucidating their respective advantages and disadvantages. The conclusion indicates that different methodologies possess distinct characteristics. POCT methods offer greater convenience and simplicity; however, they also suffer from drawbacks such as poor anti-interference capability and insufficient accuracy. Furthermore, performance varies among different POCT methods. From an accuracy standpoint, the electrochemical method demonstrates the highest accuracy, followed by flow-based, viscoelastic, and finally impedance and optical methods. Regarding user convenience, most methodologies (platforms) require fingertip blood sampling. Except for the optical method, which requires quantitative blood sampling, other methodologies do not impose specific requirements in this aspect.

Keywords: PT/INR; POCT; Electrochemical; Flow-based; Viscoelastic
Published: November 11, 2025
DOI: 10.54254/2753-8818/2025.AU29335
Volume: TNS Vol.147
pp. 226-233
Download PDF

References

  1. Davie E. W. Ratnoff O. D. (1964).  Waterfall Sequence for Intrinsic Blood Clotting. Science, 145(3638), 1310–1312.
  2. Nigel Mackman,   Rachel E. Tilley,   Nigel S. Key. (2007) Role of the extrinsic pathway of blood coagulation in hemostasis and thrombosis, Arterioscler Thromb Vasc Biol. Aug; 27(8): 1687-93.
  3. Steven P. Grover, Nigel Mackman. (2019) Intrinsic Pathway of Coagulation and Thrombosis, Arterioscler Thromb Vasc Biol. Mar; 39(3): 331-338.
  4. Peter B. Luppa,   Carolin Müller,   Alice Schlichtiger,   Harald Schlebusch. (2011) Point-of-care testing (POCT): Current techniques and future perspectives. Trends Analyt Chem. Mar 21; 30(6): 887–898.
  5. Monica Mir, Mark Vreeke, Ioanis Katakis. (2006) Different strategies to develop an electrochemical thrombin aptasensor. Electrochemistry Communications, (8): 505–511.
  6. Volker Unkrig, Michael Marquant, Fritz Hindelang, et al. (2003) Electrochemical sensior for determine blood clotting, correspondingsystem for measuring blood clotting and method for determining blood clotting, Aug.7, Pub.No.: US 2003/0146113 A1.
  7. Helen Berney, J.J. O’ Riordan. (2008) Analog Dialogure 42-08. Impedance Measurement Monitors Blood Coagulation, August.
  8. iline Microsystems, S.L., Donostia (ES). (2015) Mehod for Fluid Clotting Time Determineation on a Microfluidic Device. Dec.15, United States Patent, Pub.No.: US, 9213036B2.
  9. Edison, International Technidyne Corporation. (1994) Blood Coagulation Time Test Apparatus and Method. International Patent Classfication, 23 June, WO 94/14050.
  10. Ye-Seul Kim, M.D.1, et al. (2023) Comparison of the International Normalized Ratio Between a Point-of-Care Test and a Conventional Laboratory Test: the Latter Performs Better in Assessing Warfarin-induced Changes in Coagulation Factors. Ann Lab Med, 43: 337-344
  11. Bushra Moiz, et al. (2018) Prospective Comparison of Point-of-Care Device and Standard Analyzer for Monitoring of International Normalized Ratio in Outpatient Oral Anticoagulant Clinic, Clin Appl Thromb Hemost, Jan 28; 24(7): 1153-1158.
  12. Nurul Najmi Muhammad, et al. (2019) Comparative Evaluation of International Normalized Ratio (INR) Monitoring Between Point-Of-Care (POC) and Laboratory-Based Testing Methods in Patients Receiving Warfarin Therapy in Sultanah Nur Zahirah Hospital, pharmacy research reports, Volume 2.
  13. J. A. Peña, et al. (2012) Evaluation of the i-STAT point-of-care capillary whole blood prothrombin time and international normalized ratio: comparison to the Tcoag MDAII coagulation analyzer in the central laboratory, Clin Chim Acta.  Jun 14; 413(11-12): 955-9.
  14. Uwe Taborski,   Siegmund L. Braun,   Heinz Völler. (2004) Analytical performance of the new coagulation monitoring system INRatio for the determination of INR compared with the coagulation monitor Coaguchek S and an established laboratory method, J Thromb Thrombolysis. Oct; 18(2): 107-7.
  15. Chikayoshi Komatsu, et al. (2020) Variability of PT-INR values measured by point of care devices INRatio/INRatio 2 and CoaguChek XS and standard laboratory method. A cross-sectional study. Clinical Research. 17 August: 01~07.
  16. J. Joubert,   M. C. van Zyl,   J. Raubenheimer. (2018) Performance evaluation of the microINR®  point-of-care INR-testing system, Int J Lab Hematol. Apr; 40(2): 115-122.
  17. Alan K. Jacobson,   Jack C. Rosenfeld,   Robert R. Orr. (2021) Performance of the microINR Point-of-Care System Used by Self-Testing Patients: A Multicenter Clinical Trial, Majed A Refaai,   TH Open, Dec 30; 5(4): e577-e584.
  18. Robert D. McBane, II, MD, Carlee O’Connor, DNP, RN, et al. (2024) CoaguChek and Coag-Sense PT2 Meter Point of Care INR Device Validation. Original Article, Volume 99, Issue 7, pp.1091-1100.
  19. J. NIEDERDOCKL, C.E. DEMPFLE. (2016) Point-of-care PT and aPTT in patients with suspected deficiencies of coagulation factors. Int. Jnl. Lab. Hem. 38, 426–434.
  20. A. J. Carter,   K. Hicks,   A. W. Heldman. (1996) Clinical evaluation of a microsample coagulation analyzer, and comparison with existing techniques. Cathet Cardiovasc Diagn.  Sep; 39(1): 97-102.
  21. Yuan Teng, Shujie Yan, Gang Liu. (2022) An Agreement Study Between Point-of-Care and laboratory Activated Partial Thromboplastin Time for Anticoagulation Monitoring During Extracorporeal Membrane Oxygenation. Frontiers in Medicine, June, Volume 9, 1-8.
  22. Ashley Connelly, Rob Welner. (2019) Utility of fibrinogen in the coagulation screen. British Journal of Haematology, 186, e117–e162.