Beyond Average Accuracy: Motion-Related Failure Patterns in Wrist-Worn Photoplethysmographic Heart-Rate Monitoring
Abstract
Wrist-worn photoplethysmography (PPG) enables unobtrusive heart rate monitoring, but motion-related errors tend to cluster under specific conditions and are often masked by aggregate accuracy metrics. This study characterized those failure patterns and assessed whether wrist-sensor features could identify unreliable estimates in previously unseen participants. PPG-DaLiA recordings from 15 participants, acquired with an Empatica E4 and synchronized chest electrocardiography during daily-life activities, were segmented into 8-s windows with a 2-s step. Heart rate was estimated using a transparent dominant-frequency PPG method. A large error was defined a priori as an absolute deviation of more than 10 beats per minute (bpm) relative to the ECG-derived reference. Motion intensity, acceleration-band power, PPG-acceleration spectral overlap, peak dominance and spectral entropy were used as reliability predictors. Of 47,182 non-transition windows, 14,622 (31.0%) met the large-error criterion. Failure prevalence differed substantially by activity, from 5.8% during sitting to 63.8% during stair climbing. In leave-one-subject-out validation, a motion-only model achieved an AUROC of 0.743, wrist-signal features increased AUROC to 0.851, and activity context increased it to 0.864. Wrist PPG reliability is therefore condition dependent, and an interpretable signal-based reliability flag may complement point heart-rate estimates. External prospective validation is required before clinical or consumer deployment.
References
- Reiss, A., Indlekofer, I., Schmidt, P., & Van Laerhoven, K. (2019). Deep PPG: Large-scale heart rate estimation with convolutional neural networks.Sensors (Basel), 19(14), 3079. doi:10.3390/s19143079.
- Bent, B., Goldfarb, M., Kibbe, W. A., & Dunn, J. P. (2020). Investigating sources of inaccuracy in wearable optical heart rate sensors.NPJ Digit Med, 3, 18. doi:10.1038/s41746-020-0226-6.
- Shcherbina, A., Mattsson, C. M., Waggott, D., Salisbury, H., Christle, J. W., Hastie, T., et al. (2017). Accuracy in wrist-worn, sensor-based measurements of heart rate and energy expenditure in a diverse cohort.J Pers Med, 7(2), 3. doi:10.3390/jpm7020003.
- Zhang, Z., Pi, Z., & Liu, B. (2015). TROIKA: A general framework for heart rate monitoring using wrist-type photoplethysmographic signals during intensive physical exercise.IEEE Trans Biomed Eng, 62(2), 522-531. doi:10.1109/TBME.2014.2359372.
- Zhang, Z. (2015). Photoplethysmography-based heart rate monitoring in physical activities via joint sparse spectrum reconstruction.IEEE Trans Biomed Eng, 62(8), 1902-1910. doi:10.1109/TBME.2015.2406331.
- Salehizadeh, S., Dao, D., Bolkhovsky, J., Cho, C., Mendelson, Y., & Chon, K. H. (2016). A novel time-varying spectral filtering algorithm for reconstruction of motion artifact corrupted heart rate signals during intense physical activities using a wearable photoplethysmogram sensor.Sensors (Basel), 16(1), 10. doi:10.3390/s16010010.
- Biswas, D., Everson, L. R., Liu, M., Panwar, M., Verhoef, B. E., Patki, S., et al. (2019). CorNET: Deep learning framework for PPG-based heart rate estimation and biometric identification in ambulant environment.IEEE Trans Biomed Circuits Syst, 13(2), 282-291. doi:10.1109/TBCAS.2019.2892297.
- Crepaldi, S. C., Wang, J., Matsumoto, F., Takeuchi, H., Watanabe, T., & Yamamoto, Y. (2025). A stability- and aggregation-based method for heart rate estimation using photoplethysmographic signals during physical activity.Sensors (Basel), 25(14), 4315. doi:10.3390/s25144315.
- Ding, C., Guo, Z., Chen, Z., Lee, R. J., Rudin, C., & Hu, X. (2024). SiamQuality: A ConvNet-based foundation model for photoplethysmography signals.Physiol Meas, 45(8). doi:10.1088/1361-6579/ad6747.
- Charlton, P. H., Argüello-Prada, E. J., Mant, J., & Kyriacou, P. A. (2025). The MSPTDfast photoplethysmography beat detection algorithm: Design, benchmarking, and open-source distribution.Physiol Meas, 46(3). doi:10.1088/1361-6579/adb89e.
- Wilkosz, M., & Szczęsna, A. (2021). Multi-headed Conv-LSTM network for heart rate estimation during daily living activities.Sensors (Basel), 21(15), 5212. https://doi.org/10.3390/s21155212
- Nelson, B. W., & Allen, N. B. (2019). Accuracy of consumer wearable heart rate measurement during an ecologically valid 24-hour period: Intraindividual validation study.JMIR Mhealth Uhealth, 7(3), e10828. https://doi.org/10.2196/10828
- Elgendi, M. (2012). On the analysis of fingertip photoplethysmogram signals.Curr Cardiol Rev, 8(1), 14–25. https://doi.org/10.2174/157340312801784038
- UCI Machine Learning Repository. (2019).PPG-DaLiA[dataset]. https://archive.ics.uci.edu/dataset/495/ppg+dalia. https://doi.org/10.24432/C53890